Scrub Typhus RAPID

Configuration

  • 30 tests/kit box
  • 2 bottles of sample diluent
  • 2 explanatory sheets
  • 2 test procedure sheets

product-image

Benefits & effects

This Scrub Typhus Rapid provides not only rapid diagnosis of Scrub Typhus (Tsutsugamushi disease) among AFI but also sensitive detection of each IgM and IgG antibody simultaneously. Furthermore, this kit is easy to handle and provides rapid interpretation of the result with no equipment. In addition, this kit is useful for the follow up study and the estimation of reinfection and residual antibody.

Product Information

“ImmuneMed Scrub Typhus Rapid” tests the antibodies against Orientia tsutsugamushi in the patient’s plasma, serum or blood using Lateral Flow ImmunoChromatographic Assay.

After applying diluted serum, plasma or blood onto the hole of testing device, for IgM, patient`s sample first react with gold conjugated anti-human IgM antibody and for IgG, with gold conjugated protein A. After that, each reactant diffuses on the membrane to the recombinant antigens marked as test lines. If positive, the test line turns to red color because antigen-antibody-gold conjugate complex is formed on the test lines.
스크럽 - 발진티푸스

AFI(Acute Febrile Illness) urgently requires accurate differential early diagnosis and proper therapy against different causative microorganisms with similar clinical symptoms.

Features

1. Testing Device

One round hole for sample applying on the white plastic cassette and two detection windows for IgM and IgG. There are 2 different markings, such as C for control line and T for Scrub Typhus (Tsutsugamushi disease) by Orientia tsutsugamushi on each detection window.

2. Sample Diluents

Colorless liquid in the semi-transparent plastic bottle.

3. Purpose

Qualitative test for the antibodies against Orientia tsutsugamushi in the patient's serum, plasma or whole blood.
featured_images

Materials and Methods

1. Specimen

  • Use serum, plasma or whole blood as specimen.
  • Collected blood containing anti-coagulant should be used immediately or the blood kept in 2~8 ℃ for 3 days is possible to use.
  • Separated serum and plasma could be used until one month if kept in 2~8 ℃ or 1 year if kept frozen.

2. Testing Method

  • If each specimen and reagent is kept frozen, place it for 30 min at room temperature before use.
  • After pulling out the testing cassette from the closed pouch, place it to the flat position.
  • In case of serum or plasma, 3 ㎕ is diluted about 100 fold with 300 ㎕ sample diluent. If using whole blood, take 6 ㎕ and it is diluted about 50 fold with 300 ㎕ sample diluent. After that, apply 300 ㎕ of this diluted sample to the hole.
    * Or when 3 ㎕ of serum or plasma, or 6 ㎕ of whole blood are directly applied to the sample pad in the hole, immediately add 7 drops of sample diluent provided.
  • Wait for 15 min until the red line appears on the control line. Read and interpret the result exactly after 15 min from applying the diluted sample. Ignore the red line appeared after 15 min for interpretation.

3. Interpretation

  • Positive: Red line must appear on each C line of IgM and IgG in the detection windows and one or two red lines should appear on the test line, T in the windows (Fig 1. a,b,c).
  • Negative: Red lines only appear on both control lines of IgM and IgG in the detection windows and no red line should appear on the test line, T (Fig 1. d).
  • Invalid: If red line does not appear from at least one out of 2 C lines, this result is interpreted as invalid, and do test again (Fig 1. e).

4. Quality Control

  • Red lines should appear on both control lines for each testing.
  • If the red line does not appear from at least one out of two C lines, re-test is recommended because the directions may not have been followed correctly or the testing device may have deteriorated.

Precautions

  • In vitro diagnostic use only
  • Due to deterioration in quality when testing device is exposed to moisture in storage, pull out the device from the closed pouch right before use and use it within 30 min.
  • Do not touch the membrane in the detection window of this kit by hands directly, or it affects the test result.
  • Avoid moisture, direct sunlight and heat, and keep it in room temperature.
  • Do not use the kit after expiration date.
  • If the testing result of test line is positive, It means the antibody positive against Orientia tsutsugamushi. This kit should be used as the first line diagnosis only. The confirmation test would be necessary if positive, and the final diagnosis should be made by doctor after analysing this test result and other clinical data in conjunction with inquiries.
  • Especially in endemic area, when the kit shows faint line on test (T) line, and the clinical symptom is not remarkable, it may not be Scrub Typhus (Tsutsugamushi disease). So continuous monitoring is necessary for 1 to 7 days. During this period, please take the new blood and test again, if red color on T line increases, the possibility of Scrub Typhus (Tsutsugamushi disease) also increases. However, if the line is faint although the symptom is severe and suspicious, please send for indirect immunofluorescent antibody test (IFA) immediately.
  • Precautions of specimen handling and storage
    • Contaminated or heated specimen should not be used for testing.
    • Serious haemolysis or microbial contaminated specimen may cause inaccurate result.
    • Be careful with handling specimens which may impregnate dangerous sources of unknown viral or bacterial infection. Use a disposable glove when handle the specimen and wash your hands after handling.
    • Solid and liquid waste from experiments should be discarded after high-pressure autoclaving for 15 min at 121℃.